Customization: | Available |
---|---|
Power Supply: | 220V |
Certification: | CE, TUV, ISO |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Model | ST-80ZA | ST-80ZB | ST-80ZC | ST-80ZD | ||
Nominal volume | 80L | |||||
Inner boxsize | 400mm×400mm×500mm(W×H×D) | |||||
Dimensions | 650mm×1620mm×1135mm(W×H×D) | |||||
Technical Parameter | Temperature range | -20℃ ~ +150℃ | -40℃ ~ +150℃ | -60℃ ~ +150℃ | -70℃ ~ +150℃ | |
Control accuracy | Temperature resolution: 0.01°C Humidity resolution: 0.1%R.H Temperature deviation: ±0.5°C Humidity deviation: ±2%R.H Temperature uniformity:<2°C Humidity uniformity: ±3%R.H Temperature fluctuation: ±0.5°C Humidity fluctuation: ± 2%R.H | |||||
Temperature rise and fall rate | Heating time:-20°C→+150°C3°C/minaverage | Heating time:-40°C→+150°C3°C/minaverage | Heating time:-60°C→+150°C3°C/minaverage | Heating time:-70°C→+150°C3°C/minaverage | ||
Coolingtime:150°C→-20°Caverage1°C/minthroughout the process | Coolingtime:150°C→-40°Caverage1°C/minthroughout the process | Coolingtime:150°C→-60°Caverage1°C/minthroughout the process | Coolingtime:150°C→-70°Caverage1°C/minthroughout the process | |||
refrigeration system | compressor | Nidecfully hermetic compressors | ||||
Refrigerant | Environmentally friendly refrigerant R404A | |||||
Refrigeration method | Single-stage refrigeration | |||||
blower | Centrifugal fan with forced circulation | |||||
evaporator | Finned tubular heat exchanger | |||||
Condensationmethod | Air-cooled | |||||
condenser | Air-cooledfincondenser to cool the fan for heat dissipation | |||||
Filter driers | Absorb the residual moisture and acidic substances in the refrigerant of the refrigeration system, and filter out the solidimpurity particles and copperchipsin the system, so as to protect the normal operation of the expansion valve and capillary tube fromice plugging and dirty blockage | |||||
Humidification system | Stainless steel humidifier | |||||
Heating system | Nichrome stainless steel heater | |||||
water supply system | Front-mounted water supply tank, fully automatic water supply system |
control system | Communication interface | 1. R232/485 communication 2.It can display curvesandobtain data 3.It can be used as a monitoring and remote control system |
How to set it up | Chinese/English interface, touch input | |
How it works | Procedural/Valuation Method | |
Set the range | The maximum temperature range is5°C | |
Displaysthe resolution | Temperature:0.01°C; Time:0.1min; | |
Power-off memory function | Thepower-off recovery mode can be set as: hot start/cold start/stop | |
input | Platinum resistance | |
Storage features | The standard configuration recordsthe test data curve, can upload and download the data, and can expand the storage capacity | |
The environment in which the software is used | IBM PCcompatible machine,PII. or aboveCPU,128Mor above memory, Chinese SimplifiedWindows 2000or Chinese Simplified WindowsXPoperating system | |
Security measures | The overall equipment is over-temperature, the fan is overheated and the overall equipment is under-phase/reversed-phase, the refrigeration system is overloaded, the refrigeration unit is overpressurized, the overall equipment is timed, the leakage is leakage, the operation indication, and the automatic shutdown after the fault alarm is protected | |
structure | Chassis enclosure | High-qualitycold-rolled steel plate electrostatic spraying |
Chassis inner wall | SUS304# stainless steel plate | |
Thermal insulation | 100mm polyurethane rigid foam(or optional:high-density glass fiber woolabove 150°C). | |
Standard configuration | φ50 test hole, illuminator, double-layer vacuumtemperedglass observation window | |
weight | About180kg | |
current | 25A | |
Rated power | 5.0kw | |
power supply | Single-phase, 220V |