Customization: | Available |
---|---|
Power Supply: | 220V/380V |
Certification: | CE, TUV, ISO |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
model | ST-80LD | ST-100LD | ST-1000LD |
Nominal internal volume | 80L | 100L | 1000L |
Inner box size | 400mm×500mm×400mm(W×H×D) | 500mm×500mm×400mm (W× H × D) | 1000mm×1000mm×1000mm(W×H×D) |
External dimensions | 850mm×1420mm×955mm(W×H×D) | 950mm×1420mm×955mm (W× H × D) | 1450mm×1940mm×1515mm(W×H×D) |
Technical parameters | temperature range | -70℃ ~ + 150℃ | |
Temperature and Humidity Range Table | |||
control accuracy | Temperature resolution: 0.01 ℃ Temperature deviation: ± 0.5 ℃ Temperature uniformity: ≤ 2 ℃ Temperature fluctuation: ± 0.5 ℃ | Humidity resolution: 0.1% R.H Humidity deviation: ± 2% R.H Humidity uniformity: ± 3% R.H Humidity fluctuation: ± 2% R.H | |
Heating and cooling rate | Heating time: -70 ℃ →+150 ℃, with an average of ≥ 3 ℃/min throughout the entire process Cooling time: 150 ℃ → -70 ℃, with an average of ≥ 1 ℃/min throughout the entire process | ||
Refrigeration system | compressor | Nippon Electric's fully enclosed compressor | |
Refrigerant | Environmentally friendly refrigerant R404A/R23 | ||
Cooling method | Mechanical binary refrigeration method | ||
Blower fan | Centrifugal fan with forced circulation | ||
evaporator | Finned tube heat exchanger | ||
Condensation method | Air cooling | ||
condenser | Stainless steel brazed plate heat exchanger | ||
Dry filter | Absorb residual moisture and acidic substances in the refrigerant of the refrigeration system, and filter out solid impurity particles in the system Copper shavings, etc., protect the normal operation of the expansion valve and capillary tube from ice blockage and dirty blockage | ||
Humidification system | Stainless steel humidifier | ||
heating system | Nickel chromium alloy stainless steel heater | ||
Water supply system | Front mounted water supply tank, fully automatic water supply system |
Control system | controller | Lugong controller |
Screen display function | 1. Temperature and humidity settings (SV) and actual (PV) values are directly displayed. 2. Program execution number, segment number, remaining time, cycle number, and running time can be displayed. 3. Program editing and graphical curve display. 4. With a separate program editing screen, each page can input 4 segment temperature and humidity values and time. 5. Fixed point or program action status display 6. Language conversion can switch between Chinese/English 7.7-inch true color display screen | |
Program capacity and control function | 1. Available program quantity: up to 120 groups, 1 program can be composed of 1-99 segments, capacity: 1200 segments 2. Repeatable command execution: Each command can be executed up to 999 times 3. Programs can be linked for use; The production of the program adopts a conversational style with simple operation; 4. It has 9 sets of PID parameter settings 5. The program has skip and hold functions during execution 6. It has power-off program memory, automatically starts and continues to execute program functions after power is restored | |
Control system | Communication interface | 1. R232/485 communication 2. Can display curves and obtain data 3. Can be used as a monitoring and remote control system |
Setting method | Chinese/English interface, touch input | |
Operation mode | Program mode/fixed value mode | |
Set Range | Maximum temperature range with upper and lower limits of 5 ℃ | |
Display resolution | Temperature: 0.01 ℃; Time: 0.1 minutes; | |
Power off memory function | The power-off recovery mode can be set to: hot start/cold start/stop | |
input | rtd | |
Storage function | Standard configuration record test data curve, can upload and download data, expandable storage capacity | |
Software usage environment | IBM PC compatible machine, with a P Ⅱ or higher CPU, 128MB or higher memory, Simplified Chinese Windows 2000 or Simplified Chinese Chinese Windows XP operating system | |
Security protection measures | Overall equipment overheating, fan overheating protection, overall equipment under phase/reverse phase, refrigeration system overload, refrigeration unit overpressure, overall equipment timing, leakage, operation indication, automatic shutdown after fault alarm, and other protections | |
structure | Chassis shell | High quality cold-rolled steel plate electrostatic spraying (1.2mm thick) |
Inner wall of chassis | SUS304 # stainless steel plate (1.2mm thick) | |
Thermal insulation material | 100mm polyurethane rigid foam+100mm high-density glass fiber cotton | |
Standard configuration | 50 diameter test hole, lighting fixture, double-layer vacuum strengthened glass observation window | |
weight | 195kg | |
electric current | 30A | |
Rated power | 6.5kw |